Media-fill simulation tests in manual and robotic aseptic preparation of injection solutions in syringes
- PMID: 25549919
- DOI: 10.1177/1078155214565123
Media-fill simulation tests in manual and robotic aseptic preparation of injection solutions in syringes
Abstract
Objective: The purpose of this study was to evaluate the contamination rate of media-fill products either prepared automated with a robotic system (APOTECAchemo™) or prepared manually at cytotoxic workbenches in the same cleanroom environment and by experienced operators. Media fills were completed by microbiological environmental control in the critical zones and used to validate the cleaning and disinfection procedures of the robotic system.
Methods: The aseptic preparation of patient individual ready-to-use injection solutions was simulated by using double concentrated tryptic soy broth as growth medium, water for injection and plastic syringes as primary packaging materials. Media fills were either prepared automated (500 units) in the robot or manually (500 units) in cytotoxic workbenches in the same cleanroom over a period of 18 working days. The test solutions were incubated at room temperature (22℃) over 4 weeks. Products were visually inspected for turbidity after a 2-week and 4-week period. Following incubation, growth promotion tests were performed with Staphylococcus epidermidis. During the media-fill procedures, passive air monitoring was performed with settle plates and surface monitoring with contact plates on predefined locations as well as fingerprints. The plates got incubated for 5-7 days at room temperature, followed by 2-3 days at 30-35℃ and the colony forming units (cfu) counted after both periods. The robot was cleaned and disinfected according to the established standard operating procedure on two working days prior to the media-fill session, while on six other working days only six critical components were sanitized at the end of the media-fill sessions. Every day UV irradiation was operated for 4 h after finishing work.
Results: None of the 1000 media-fill products prepared in the two different settings showed turbidity after the incubation period thereby indicating no contamination with microorganisms. All products remained uniform, clear, and light-amber solutions. In addition, the reliability of the nutrient medium and the process was demonstrated by positive growth promotion tests with S. epidermidis. During automated preparation the recommended limits < 1 cfu per settle/contact plate set for cleanroom Grade A zones were not succeeded in the carousel and working area, but in the loading area of the robot. During manual preparation, the number of cfus detected on settle/contact plates inside the workbenches lay far below the limits. The number of cfus detected on fingertips succeeded several times the limit during manual preparation but not during automated preparation. There was no difference in the microbial contamination rate depending on the extent of cleaning and disinfection of the robot.
Conclusion: Extensive media-fill tests simulating manual and automated preparation of ready-to-use cytotoxic injection solutions revealed the same level of sterility for both procedures. The results of supplemental environmental controls confirmed that the aseptic procedures are well controlled. As there was no difference in the microbial contamination rates of the media preparations depending on the extent of cleaning and disinfection of the robot, the results were used to adapt the respective standard operating procedures.
Keywords: Media fill; aseptic preparation; cleaning validation; environmental monitoring; simulation.
© The Author(s) 2014.
Similar articles
-
Implementation and microbiological stability of dose-banded ganciclovir infusion bags prepared in series by a robotic system.Eur J Hosp Pharm. 2020 Jul;27(4):209-215. doi: 10.1136/ejhpharm-2018-001745. Epub 2018 Nov 26. Eur J Hosp Pharm. 2020. PMID: 32587079 Free PMC article.
-
Microbiological performance of a robotic system for aseptic compounding of cytostatic drugs.Eur J Pharm Sci. 2019 Mar 15;130:181-185. doi: 10.1016/j.ejps.2019.01.034. Epub 2019 Jan 31. Eur J Pharm Sci. 2019. PMID: 30710619
-
Aseptic simulation test challenged with microorganisms for validation of pharmacy operators.Am J Health Syst Pharm. 2012 Jul 15;69(14):1218-24. doi: 10.2146/ajhp110295. Am J Health Syst Pharm. 2012. PMID: 22761076
-
Industrial manufacture of parenteral products in The Netherlands. A survey of eight years of media fills and sterility testing.PDA J Pharm Sci Technol. 1998 Jul-Aug;52(4):159-64. PDA J Pharm Sci Technol. 1998. PMID: 9752710 Review.
-
Microbial diversity in pharmaceutical product recalls and environments.PDA J Pharm Sci Technol. 2007 Sep-Oct;61(5):383-99. PDA J Pharm Sci Technol. 2007. PMID: 18047177 Review.
Cited by
-
Validation of the Media Fill Method for Solid Tissue Processing in Good Manufacturing Practice-Compliant Cell Production.Methods Mol Biol. 2021;2286:67-71. doi: 10.1007/7651_2020_331. Methods Mol Biol. 2021. PMID: 33349901
-
Microbiological validation of a robot for the sterile compounding of injectable non-hazardous medications in a hospital environment.Eur J Hosp Pharm. 2020 Mar;27(e1):e63-e68. doi: 10.1136/ejhpharm-2018-001757. Epub 2019 Feb 4. Eur J Hosp Pharm. 2020. PMID: 32296508 Free PMC article.
-
Implementation and microbiological stability of dose-banded ganciclovir infusion bags prepared in series by a robotic system.Eur J Hosp Pharm. 2020 Jul;27(4):209-215. doi: 10.1136/ejhpharm-2018-001745. Epub 2018 Nov 26. Eur J Hosp Pharm. 2020. PMID: 32587079 Free PMC article.
-
Implementation and validation of an in-house combined fluorescein/media-fill test to qualify radiopharmacy operators.EJNMMI Radiopharm Chem. 2021 Jan 7;6(1):2. doi: 10.1186/s41181-020-00117-6. EJNMMI Radiopharm Chem. 2021. PMID: 33411035 Free PMC article.
-
Analysis of production time and capacity for manual and robotic compounding scenarios for parenteral hazardous drugs.Eur J Hosp Pharm. 2024 Jun 21;31(4):352-357. doi: 10.1136/ejhpharm-2022-003576. Eur J Hosp Pharm. 2024. PMID: 36792350
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials