Total erythrocyte destruction time: the new index for the hemolytic performance of rotary blood pumps
- PMID: 8572954
- DOI: 10.1111/j.1525-1594.1995.tb02383.x
Total erythrocyte destruction time: the new index for the hemolytic performance of rotary blood pumps
Abstract
To date an index of hemolysis is the only parameter to evaluate in vitro the hemolytic properties of blood pumps. However, this index does not reflect the sublethal conditions of erythrocytes. In this paper, a new concept, the total destruction of erythrocytes, to indicate traumatic features on erythrocytes is proposed. Five paired long duration in vitro hemolysis tests were conducted using the BioMedicus pump; BP80, and the Nikkiso pump, HMS12. The pumps ran until either a sudden change in blood color to dark red was noted or the pumps ran for over 24 h. At a certain point, the plasma-free hemoglobin levels suddenly and dramatically increased in each experiment except one Nikkiso case. At a certain point in time, the destruction of most erythrocytes suddenly and simultaneously occurred. We call this phenomenon total destruction of erythrocytes and define the pumping duration required to raise the level of the plasma-free hemoglobin to 50% of the total hemoglobin as total destruction time. Although there is no statistical difference in the normalized index of hemolysis between the two pumps, the total destruction time of the BioMedicus pump is significantly shorter than that of the Nikkiso pump. It appears that an accumulation of sublethal damage on erythrocytes resulted in the decrease of the hemolytic threshold of the entire erythrocyte and also resulted in a sudden entire destruction of erythrocytes.
Similar articles
-
New method of evaluating sublethal damage to erythrocytes by blood pumps.Artif Organs. 1996 Jun;20(6):568-71. Artif Organs. 1996. PMID: 8817957
-
Reconsideration of total erythrocyte destruction phenomenon.Artif Organs. 1997 Jul;21(7):704-9. doi: 10.1111/j.1525-1594.1997.tb03727.x. Artif Organs. 1997. PMID: 9212943
-
New Method of Evaluating Sublethal Damage to Erythrocytes by Blood Pumps.Artif Organs. 1996 May;20(5):568-571. doi: 10.1111/j.1525-1594.1996.tb04482.x. Artif Organs. 1996. PMID: 28868719
-
Biomaterials for rotary blood pumps.Artif Organs. 1995 Jul;19(7):603-7. doi: 10.1111/j.1525-1594.1995.tb02388.x. Artif Organs. 1995. PMID: 8572959 Review.
-
Mechanical blood trauma in assisted circulation: sublethal RBC damage preceding hemolysis.Int J Artif Organs. 2016 Jun 15;39(4):150-9. doi: 10.5301/ijao.5000478. Epub 2016 Mar 30. Int J Artif Organs. 2016. PMID: 27034320 Free PMC article. Review.
Cited by
-
Technical-Induced Hemolysis in Patients with Respiratory Failure Supported with Veno-Venous ECMO - Prevalence and Risk Factors.PLoS One. 2015 Nov 25;10(11):e0143527. doi: 10.1371/journal.pone.0143527. eCollection 2015. PLoS One. 2015. PMID: 26606144 Free PMC article.
-
Risk factors associated with 30-day mortality for out-of-center ECMO support: experience from the newly launched ECMO retrieval service.J Artif Organs. 2019 Jun;22(2):110-117. doi: 10.1007/s10047-019-01092-9. Epub 2019 Jan 23. J Artif Organs. 2019. PMID: 30673894
-
Significance of hemolysis on extracorporeal life support after cardiac surgery in children.Pediatr Nephrol. 2009 Mar;24(3):589-95. doi: 10.1007/s00467-008-1047-z. Epub 2008 Nov 12. Pediatr Nephrol. 2009. PMID: 19002722
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous