Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome
- PMID: 17440422
- DOI: 10.1097/01.CCM.0000266586.04676.55
Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome
Abstract
Objective: a) Characterize how ventilator and patient variables affect tidal volume during high-frequency oscillatory ventilation; and b) measure tidal volumes in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation.
Design: Observational study.
Setting: Research laboratory and medical intensive care unit.
Patients: Test lung and patients with acute respiratory distress syndrome.
Interventions: Using a previously validated hot wire anemometer placed in series with a Sensormedics 3100B high-frequency ventilator, an endotracheal tube, and a test lung, tidal volume was measured at different combinations of frequency (4, 6, 8, 10, and 12 Hz), pressure amplitude (50, 60, 70, 80, and 90 cm H2O), mean airway pressure (20, 30, and 40 cm H2O), test lung compliance (10, 30, and 50 mL/cm H2O), endotracheal tube internal diameter (6, 7, and 8 mm), bias flow (20, 30, and 40 L/min), and inspiratory/expiratory ratio (1:2 and 1:1). In patients, tidal volume was measured at baseline ventilator settings and at baseline frequency +/-2 Hz and baseline pressure amplitude +/-10 cm H2O.
Measurements and main results: Measured tidal volumes were 23-225 mL during high-frequency oscillatory ventilation of the test lung. A 2-Hz increase in frequency and a 10-cm H2O increase in pressure amplitude caused a 21.3% +/- 4.1% decrease and 21.4% +/- 3.4% increase in tidal volume, respectively. Decreasing endotracheal tube internal diameter from 8 mm to 7 mm and from 7 mm to 6 mm caused a 15.3% +/- 1.7% and 18.9% +/- 2.1% reduction in tidal volume, respectively. Increasing bias flow from 20 L/min to 30 L/min increased tidal volume by 11.2% +/- 3.9%. Further increases in bias flow, changes in compliance, and changes in mean airway pressure had little effect. Tidal volumes measured in acute respiratory distress syndrome patients were 44-210 mL. A 2-Hz increase in frequency was associated with a 23.1% +/- 6.3% decrease in tidal volume. In contrast to the test lung data, a 10-cm H2O increase in pressure amplitude resulted in only a 5.6% +/- 4.5% increase in tidal volume.
Conclusions: Tidal volumes are not uniformly small during high-frequency oscillatory ventilation. The primary determinant of tidal volume in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation with the Sensormedics 3100B is frequency. Test lung findings suggest that endotracheal tube internal diameter is also an important determinant of tidal volume.
Comment in
-
High-frequency oscillation: how high should we go?Crit Care Med. 2007 Jun;35(6):1623-4. doi: 10.1097/01.CCM.0000267655.07864.26. Crit Care Med. 2007. PMID: 17522540 No abstract available.
Similar articles
-
Four methods of measuring tidal volume during high-frequency oscillatory ventilation.Crit Care Med. 2006 Mar;34(3):751-7. doi: 10.1097/01.CCM.0000201400.63304.41. Crit Care Med. 2006. PMID: 16505661
-
In vitro performance comparison of the Sensormedics 3100A and B high-frequency oscillatory ventilators.Pediatr Crit Care Med. 2011 Jul;12(4):e176-80. doi: 10.1097/PCC.0b013e3181fe3028. Pediatr Crit Care Med. 2011. PMID: 21037502
-
Feasibility of very high-frequency ventilation in adults with acute respiratory distress syndrome.Crit Care Med. 2008 Apr;36(4):1043-8. doi: 10.1097/01.CCM.0b013e318168fcab. Crit Care Med. 2008. PMID: 18379227
-
A protocol for high-frequency oscillatory ventilation in adults: results from a roundtable discussion.Crit Care Med. 2007 Jul;35(7):1649-54. doi: 10.1097/01.CCM.0000269026.40739.2E. Crit Care Med. 2007. PMID: 17522576
-
High-frequency oscillatory ventilation in adults with acute respiratory distress syndrome.Curr Opin Anaesthesiol. 2012 Feb;25(1):17-23. doi: 10.1097/ACO.0b013e32834ea57b. Curr Opin Anaesthesiol. 2012. PMID: 22157194 Review.
Cited by
-
Ventilation in chest trauma.J Emerg Trauma Shock. 2011 Apr;4(2):251-9. doi: 10.4103/0974-2700.82215. J Emerg Trauma Shock. 2011. PMID: 21769213 Free PMC article.
-
Novel Mechanical Strain Characterization of Ventilated ex vivo Porcine and Murine Lung using Digital Image Correlation.Front Physiol. 2020 Dec 4;11:600492. doi: 10.3389/fphys.2020.600492. eCollection 2020. Front Physiol. 2020. PMID: 33343395 Free PMC article.
-
"Low-" versus "high"-frequency oscillation and right ventricular function in ARDS. A randomized crossover study.J Intensive Care. 2018 Sep 4;6:58. doi: 10.1186/s40560-018-0327-3. eCollection 2018. J Intensive Care. 2018. PMID: 30202530 Free PMC article.
-
The design of future pediatric mechanical ventilation trials for acute lung injury.Am J Respir Crit Care Med. 2010 Dec 15;182(12):1465-74. doi: 10.1164/rccm.201004-0606CI. Epub 2010 Aug 23. Am J Respir Crit Care Med. 2010. PMID: 20732987 Free PMC article. Review.
-
Comparison of the ventilation characteristics in two adult oscillators: a lung model study.Intensive Care Med Exp. 2019 Mar 12;7(1):15. doi: 10.1186/s40635-019-0229-2. Intensive Care Med Exp. 2019. PMID: 30868327 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources