Monitoring the traceability of the pH of a primary tetraborate buffer: comparison of results from primary and secondary apparatus
- PMID: 17245528
- DOI: 10.1007/s00216-007-1117-2
Monitoring the traceability of the pH of a primary tetraborate buffer: comparison of results from primary and secondary apparatus
Abstract
This paper reports evaluation of the behaviour of different combined glass electrodes applied to measurement of the pH of a primary, 0.01 mol kg(-1), tetraborate buffer. Measurements were first performed by use of a primary Harned cell (at 15, 25, and 37 degrees C); these results were then compared with those obtained for the same solution by use of three combined glass electrodes (25 degrees C) with different membranes and liquid-junction designs, calibrated by use of commercial pH-metric buffers. The pH of the same solution was also measured in terms of the molal concentration of hydrogen ions, using acid-base titration to evaluate the formal potential difference K of each cell at fixed ionic strength, I, adjusted by addition of KCl or Et4NI (tetraethylammonium iodide). The reference value from primary measurement, paH = 9.171, was slightly closer to the mean value obtained by determination of concentration, rather than that obtained by direct measurement of activity; the differences were smaller than the extended uncertainty characteristics of the secondary measurements. The importance of evaluation of the ionic strength of the solution under study is emphasised. We verified that for tetraborate buffer slight modification of the value of I used to calculate gamma (i) (the activity coefficient of a single ion) in the calculation of paH from the acidity function at zero molality of chloride can significantly affect the reference value of the calibrator tool. This is true, in general, for low values of the ionic strength, such as those considered in this work; an approximate value of I can then cause distortions along the pH traceability chain. Application of the concepts of thermodynamics to this traceability chain is discussed.
Similar articles
-
Role of the activity coefficient in the dissemination of pH: comparison of primary (Harned cell) and secondary (glass electrode) measurements on phosphate buffer considering activity and concentration scales.Anal Bioanal Chem. 2005 Sep;383(2):341-8. doi: 10.1007/s00216-005-3417-8. Epub 2005 Sep 2. Anal Bioanal Chem. 2005. PMID: 16142421
-
Acidity function p(a(H) gamma(Cl)) as a step to pH assessment.Anal Bioanal Chem. 2002 Nov;374(5):806-12. doi: 10.1007/s00216-002-1373-0. Epub 2002 Jul 4. Anal Bioanal Chem. 2002. PMID: 12434234
-
Development of a reference solution for the pH of seawater.Anal Bioanal Chem. 2007 Aug;388(8):1877-83. doi: 10.1007/s00216-007-1423-8. Epub 2007 Jul 7. Anal Bioanal Chem. 2007. PMID: 17619179
-
Measuring the pH of Nanoparticle Suspensions: Version 1.1.2010 Mar. In: National Cancer Institute’s Nanotechnology Characterization Laboratory Assay Cascade Protocols [Internet]. Bethesda (MD): National Cancer Institute (US); 2005 May 1–. NIST - NCL Joint Assay Protocol, PCC-13. 2010 Mar. In: National Cancer Institute’s Nanotechnology Characterization Laboratory Assay Cascade Protocols [Internet]. Bethesda (MD): National Cancer Institute (US); 2005 May 1–. NIST - NCL Joint Assay Protocol, PCC-13. PMID: 39013068 Free Books & Documents. Review.
-
Electrodes in clinical chemistry.Ann Clin Biochem. 1985 Sep;22 ( Pt 5):460-88. doi: 10.1177/000456328502200503. Ann Clin Biochem. 1985. PMID: 3904587 Review.
LinkOut - more resources
Full Text Sources